FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 1895682
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04488
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- August 17, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.
Description of Event or Problem · 1
IT WAS REPORTED THAT NO SENSING AND LOSS OF CAPTURE WERE OBSERVED. FLUOROSCOPY REVEALED THAT THE ATRIAL AND RV LEAD HAD ROLLED UP AROUND THE GENERATOR AND WERE DISLODGED. THE LEADS WERE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention | (B)(4) |