FDA Adverse Event Malfunction Summary report: N

SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER

MDR report key: 1895672 · Received November 10, 2010

Report

Report Number
2050012-2010-01199
Event Type
Malfunction
Date Received
November 10, 2010
Date of Event
October 12, 2010
Report Date
November 10, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K188495
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER TECHNICAL SUPPORT (CTS) HAD THE CUSTOMER CHECK THE REAGENT SYRINGE AND PROBES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 AND INITIALLY CHECKED THE ALIGNMENTS FOR THE REAGENT PROBE AND MIXER. THE FSE THEN REPLACED THE VALVE FOR THE REAGENT PROBE. ALL CONTROLS WERE RUNNING TO SPECIFICATION, BUT SPLASHING STILL OCCURRING. SERVICE INSTALLED A NEW REAGENT PROBE AND SYRINGE. THE FSE ADJUSTED THE REAGENT PROBE DRIVE BELT TENSION, WHICH RESOLVED THE SPLASHING ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT A SPLASHING ON THE REAGENT TRAY WAS OBSERVED AFTER PREVENTATIVE MAINTENANCE (PM) WAS PERFORMED ON THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER. NO OPERATOR EXPOSURE WAS NOTED.