FDA Adverse Event
Malfunction
Summary report: N
SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER
MDR report key: 1895672
·
Received November 10, 2010
Report
- Report Number
- 2050012-2010-01199
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- October 12, 2010
- Report Date
- November 10, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K188495
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER TECHNICAL SUPPORT (CTS) HAD THE CUSTOMER CHECK THE REAGENT SYRINGE AND PROBES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2010 AND INITIALLY CHECKED THE ALIGNMENTS FOR THE REAGENT PROBE AND MIXER. THE FSE THEN REPLACED THE VALVE FOR THE REAGENT PROBE. ALL CONTROLS WERE RUNNING TO SPECIFICATION, BUT SPLASHING STILL OCCURRING. SERVICE INSTALLED A NEW REAGENT PROBE AND SYRINGE. THE FSE ADJUSTED THE REAGENT PROBE DRIVE BELT TENSION, WHICH RESOLVED THE SPLASHING ISSUE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) STATING THAT A SPLASHING ON THE REAGENT TRAY WAS OBSERVED AFTER PREVENTATIVE MAINTENANCE (PM) WAS PERFORMED ON THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER. NO OPERATOR EXPOSURE WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | CX5 DELTA | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |