FDA Adverse Event
Malfunction
Summary report: N
RIATA ACTIVE FIXATION
MDR report key: 1895606
·
Received November 10, 2010
Report
- Report Number
- 2017865-2010-04347
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- August 13, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(6). NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. ANALYSIS REVEALED OUTER INSULATION ABRASION AT 10.2CM FROM THE CONNECTOR PIN, DUE TO FRICTION TO THE ICD CAN.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1581/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |