CURRENT RF DR
Report
- Report Number
- 2017865-2010-04385
- Event Type
- Injury
- Date Received
- November 10, 2010
- Date of Event
- August 3, 2010
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
THE REPORTED PREMATURE BATTERY DEPLETION WAS VERIFIED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, THE DEVICE WAS FOUND TO BE BELOW THE EXPECTED LIMITS. NO HIGH CURRENT DRAIN SOURCES WERE FOUND DURING BENCH, AUTOMATED ELECTRICAL, TEMPERATURE, AND HUMIDITY TESTING. THE BATTERY WAS SENT TO THE VENDOR FOR FURTHER ANALYSIS. AN INTERNAL BATTERY ANOMALY WAS FOUND TO CAUSE THE PREMATURE BATTERY DEPLETION.
PATIENT PRESENTED IN THE CLINIC WITH DEVICE AT ERI. LONGEVITY ESTIMATION NOTED THAT THE DEVICE WAS EXPECTED TO LAST LONGER THAN OBSERVED. BATTERY VOLTAGE TREND IS REPORTEDLY SHOWING A RAPID DEPLETION. THE PATIENT REQUESTED NOT TO PERFORM ANY PROCEDURE. THE DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CURRENT RF DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 2207-36 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention |