FDA Adverse Event Injury Summary report: N

PROSTAR

MDR report key: 18955055 · Received March 21, 2024

Report

Report Number
2024168-2024-03607
Event Type
Injury
Date Received
March 21, 2024
Date of Event
March 4, 2024
Report Date
August 6, 2024
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648000621
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO FIRE WAS NOT CONFIRMED. UNUSED STERILE DEVICES WERE RETURNED FROM THE ACCOUNT. THE STERILE UNITS WERE VISUALLY AND FUNCTIONALLY TESTED WITH NO ANOMALIES NOTED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. FACTORS THAT MAY CONTRIBUTE TO THE REPORTED FAILURE TO DEPLOY THE NEEDLES MAY INCLUDE, BUT ARE NOT LIMITED TO, CHANGE POSITION OF DEVICE DURING DEPLOYMENT, DEVICE ORIENTATION, OVER INSERTION OF THE DEVICE. THE REPORTED DIFFICULTIES AND TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE REPORTED INFORMATION AND RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D1 CORRECTION: BRAND NAME UPDATEDD2A CORRECTION: COMMON DEVICE NAME UPDATEDD3 CORRECTION: NAME, ADDRESS UPDATEDCORRECTION D4 - LOT # / D4 - UDI # UPDATED FROM UNKNOWN/08717648000621 TO: 3111441 / (01)08717648000621(17)250930(10)3111441.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PROSTAR REFERENCED IN B5 IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VESSEL CLOSURE OF AN UNSPECIFIED ACCESS SITE USING TWO PROSTAR DEVICES. REPORTEDLY, THE NEEDLE DID NOT LINE UP WITH THE OPENING FOR BOTH DEVICES. AN ADDITIONAL PROSTAR DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2555186 PROSTAR DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12322-01 3111441 08717648000621

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention