FDA Adverse Event Injury Summary report: N

TUBE, GASTROINTESTINAL (AND ACCESS)

MDR report key: 1895379 · Received November 4, 2010

Report

Report Number
3006260740-2010-00325
Event Type
Injury
Date Received
November 4, 2010
Date of Event
October 13, 2010
Report Date
October 19, 2010
Manufacturer
C. R. BARD, INC. (BASD)
Product Code
KNT
PMA / PMN Number
K915837
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT OF A BREAK IN THE RETENTION DOME IS CONFIRMED, USER RELATED. UPON RECEIPT, A PARTIAL SEMI-CIRCULAR BREAK IN THE RETENTION DOME WAS OBSERVED. A SMALL SECTION OF THE BROKEN DOME IS STILL ATTACHED TO THE FEEDING TUBE AND REVEALS A COMPLETE BOND. THE DULL AND ROUNDED VENEER IDENTIFIED AT THE DOME SITE ARE TRAITS THAT ARE INDICATIVE OF EXCESSIVE FORCE THAT WAS APPLIED DURING THE REMOVAL OF THIS DEVICE. THE LACK OF ANY ASSOCIATED DAMAGE SUGGESTS THE TEAR IN THE DOME WAS CAUSED BY STRETCHING THE DOME MATERIAL BEYOND ITS ELASTIC LIMITS DURING REMOVAL. GROSS VISUAL AND MICROSCOPIC EXAMINATIONS SHOW NO EVIDENCE OF A DEFECT OR DEFORMITY RELATED TO THE MANUFACTURING PROCESS. A CHR OF LOT #HUTA1528 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER.

Description of Event or Problem · 1

A BREAK IN THE RETENTION DOME DURING TRACTION REMOVAL. THE INDWELL TIME WAS 212 DAYS. THE DOME PORTION WAS REMOVED ENDOSCOPICALLY USING GRASPING FORCEPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TUBE, GASTROINTESTINAL (AND ACCESS) KNT C. R. BARD, INC. (BASD) HUTA1528

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention