FDA Adverse Event Injury Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE)

MDR report key: 1895325 · Received November 2, 2010

Report

Report Number
2024601-2010-00895
Event Type
Injury
Date Received
November 2, 2010
Date of Event
September 8, 2010
Report Date
September 16, 2010
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

TAPER I. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. REFLUX IS A SURGICAL/PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. ANALYSIS OF THE SHELL NOTED AN UNIDENTIFIED SHARP, CURVED OPENING WITH LEAKAGE. THE BAND TUBING AND BUCKLE WERE BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF REFLUX AS FOLLOWS: "ULCERATION, GASTRITIS, GASTROESOPHAGEAL REFLUX, HEARTBURN, GAS BLOAT, DYSPHAGIA, DEHYDRATION, CONSTIPATION AND WEIGHT REGAIN HAVE BEEN REPORTED AFTER GASTRIC RESTRICTION PROCEDURES."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED THAT THERE WAS AN EXPLANTED DEVICE IN THEIR POSSESSION THAT NEEDED TO BE RETURNED. NO ADDITIONAL INFORMATION WAS PROVIDED. FURTHER FOLLOW-UP WITH THE HEALTH PROFESSIONAL NOTED "BAND/PORT EXPLANT DUE TO REFLUX."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE) LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 42 YR Required Intervention