FDA Adverse Event
Injury
Summary report: N
PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
MDR report key: 1895240
·
Received November 8, 2010
Report
- Report Number
- 2953144-2010-02750
- Event Type
- Injury
- Date Received
- November 8, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 19, 2010
- Manufacturer
- ABBOTT VASCULAR-VASCULAR SOLUTIONS
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4) - PATIENT SELECTION - HEAVILY CALCIFIED ARTERY, (B)(4). THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A HEAVILY CALCIFIED COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, NO SUTURE WAS ATTACHED TO THE NEEDLES. THE DEVICE WAS REMOVED OVER A GUIDE WIRE, AND A SECOND PROGLIDE DEVICE WAS SUCCESSFULLY USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | MGB | ABBOTT VASCULAR-VASCULAR SOLUTIONS | NA | 930026H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |