FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1894863 · Received November 5, 2010

Report

Report Number
2531779-2010-02155
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 7, 2010
Report Date
October 7, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED THAT THE PUMP WAS NOT RESPONDING TO BOLUS BUTTON PRESSES. THE BOLUS BUTTON WAS REMOVED AND OBSERVED TO BE OFF CENTERED.

Description of Event or Problem · 1

THE PT CONTACTED ANIMAS ALLEGING THAT THE PUMP WAS NOT RESPONDING TO BUTTON PRESSES. THE PT CLAIMED THAT WHENEVER SHE PRESSED THE BUTTONS, THE PUMP WOULD VIBRATE AND SOUND AS THOUGH THE MOTOR WAS REWINDING ALTHOUGH NOTHING WOULD APPEAR ON THE SCREEN. THE PT DENIED THAT THE PUMP WAS EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR