FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1894848
·
Received November 5, 2010
Report
- Report Number
- 2531779-2010-02142
- Event Type
- Malfunction
- Date Received
- November 5, 2010
- Report Date
- October 6, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED SWOLLEN, THE UP KEY CONTACT WAS MISALIGNED, ALL KEY CONTACTS ARE INVERTED AND DO NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
THE PT CLAIMED THAT THE PUMP INTERMITTENTLY RESPONSIVE TO THE OK AND ARROW BUTTONS. THIS REPORTED ISSUE CAN IMPACT INSULIN DELIVERY. THE PUMP REPORTEDLY IS "RARELY" EXPOSED TO WATER (2X/YEAR) AND THERE ARE NO SIGN OF DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |