ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1894848 · Received November 5, 2010

Report

Report Number
2531779-2010-02142
Event Type
Malfunction
Date Received
November 5, 2010
Report Date
October 6, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED SWOLLEN, THE UP KEY CONTACT WAS MISALIGNED, ALL KEY CONTACTS ARE INVERTED AND DO NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PT CLAIMED THAT THE PUMP INTERMITTENTLY RESPONSIVE TO THE OK AND ARROW BUTTONS. THIS REPORTED ISSUE CAN IMPACT INSULIN DELIVERY. THE PUMP REPORTEDLY IS "RARELY" EXPOSED TO WATER (2X/YEAR) AND THERE ARE NO SIGN OF DAMAGE TO THE KEYPAD. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA