FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1894845 · Received November 5, 2010

Report

Report Number
2531779-2010-02156
Event Type
Malfunction
Date Received
November 5, 2010
Report Date
October 6, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE INTACT (NO LIFTING OR PEELING), BUT ALL THE KEYPAD BUTTONS WERE UNRESPONSIVE TO USER INPUT, THE UP ARROW KEY CONTACT WAS MISALIGNED, AND THE OK KEY CONTACT WAS INVERTED AND DID NOT SPRING BACK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP IS UNRESPONSIVE WHEN THE OK BUTTON IS PRESSED AND A SINGLE PRESS OF THE ARROW BUTTONS OFTEN CAUSES THE CURSOR TO SCROLL. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO WATER AND DOES NOT HAVE ANY VISIBLE DAMAGE TO THE BUTTONS. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1