FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1894844 · Received November 5, 2010

Report

Report Number
2531779-2010-02160
Event Type
Malfunction
Date Received
November 5, 2010
Report Date
October 6, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD WAS TORN OFF AT THE CONTRAST BUTTON, ALL KEYPAD BUTTONS WERE NOT RESPONDING WHEN PRESSED, THE CONTRAST KEY CONTACT WAS MISSING, THE UP ARROW KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP IS NOT RESPONDING WHEN THE UP AND DOWN ARROW BUTTONS ARE PRESSED AND THE BUTTONS FEEL FLAT. THE PT'S FATHER ALSO INDICATED THAT THERE IS A TEAR AT THE UP ARROW BUTTON. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO WATER. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 16 YR