FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1894838 · Received November 5, 2010

Report

Report Number
2531779-2010-02131
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 7, 2010
Report Date
October 7, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT THE FORCE SENSOR RESISTANCE READING WAS OUT OF CALIBRATION. IN ADDITION, VISIBLE MOISTURE INGRESS WAS OBSERVED WITHIN ON THE AREA HOUSING THE FORCE SENSOR. AN EZPRIME OPERATION WAS PERFORMED AND THE PUMP DISPENSED ALL THE INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP AND EMITTED A "NO CARTRIDGE DETECTED" WARNING. THERE WAS ONE "NO CARTRIDGE DETECTED" WARNING IN THE PUMP HISTORY. A MISALIGNED DISPLAY SCREEN WAS ALSO OBSERVED, BUT DID NOT CAUSE THE FORCE SENSOR PINS TO DISLODGE.

Description of Event or Problem · 1

THE PT REPORTED THAT THE PUMP DISPENSES INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP. THE PT CONFIRMED HIS BODY WAS NOT CONNECTED TO THE PUMP DURING PRIMING STEPS. THERE WAS NO REPORTED PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 31 YR