STRAIGHT LONG SABER ATTACHMENT

MDR report key: 1894788 · Received November 5, 2010

Report

Report Number
1811755-2010-01613
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 13, 2010
Report Date
October 13, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS RETURNED TO THE MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED, AND EXTENSIVE CORROSION AND DEBRIS WAS DISCOVERED WITHIN VARIOUS COMPONENTS, INCLUDING THE BEARINGS. THE PRESENCE OF CORROSION AND DEBRIS CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. IMPROPER OR INADEQUATE CLEANING/STERILIZATION TECHNIQUES CAN CONTRIBUTE TO THE BUILDUP OF CORROSION AND DEBRIS WITHIN THIS DEVICE. THE DRILL ATTACHMENT COULD NOT BE REPAIRED AND WAS DISCARDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EQUIPMENT EVAL CONDUCTED BY A MFR FIELD SERVICE TECHNICIAN, THE DEVICE HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRAIGHT LONG SABER ATTACHMENT DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC. ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK