FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1894725 · Received November 5, 2010

Report

Report Number
2531779-2010-02104
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 6, 2010
Report Date
October 7, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND THE EVAL REVEALED THAT THE KEYPAD APPEARED TO BE SWOLLEN. ALL KEYPAD BUTTONS WERE RESPONSIVE TO USER INPUTS. ALL KEYPAD BUTTONS CLICK AND SPRING BACK NORMALLY. THE KEYPAD WAS REMOVED AND THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. ADDITIONALLY, THE BOLUS BUTTON WAS DIFFICULT TO PUSH.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE BUTTONS WERE NOT RESPONDING ON THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA