HOMECHOICE CYCLER-REFURBISHED
Report
- Report Number
- 1423500-2010-05541
- Event Type
- Malfunction
- Date Received
- November 9, 2010
- Date of Event
- September 30, 2010
- Report Date
- October 18, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE ACTUAL DEVICE WAS RETURNED AND EVALUATED. REVIEW OF THE DEVICE LOGS REVEALED AN INCREASED INTRAPERITONEAL VOLUME (IIPV) HAD OCCURRED WHEN THE USER DRAINED OUT 3304 ML, WHICH WAS GREATER THAN THE LARGEST PRESCRIBED FILL VOLUME OF 2000 ML AND MET IIPV CRITERIA. THE ASSIGNABLE CAUSE OF THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE: INSUFFICIENT DRAIN- FALSE EMPTY DETECT AND USE ERROR INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED TOO LOW (0ML) AND USE ERROR, TIDAL UF REMOVAL SET TOO LOW (0ML). A REVIEW OF THE PREVIOUS SERVICE RECORD SHOWS THE DEVICE PASSED ALL REQUIRED TESTS AND CALIBRATIONS PRIOR TO ITS RELEASE FROM THE (B)(4). NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE ISSUE OF THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WHICH OCCURRED ON (B)(6) 2010 DURING NITE 4 ULTRAFILTRATION, DRAIN VOLUME 3304ML. THIS EVENT MEETS THE OVERFILL CRITERIA
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE CYCLER-REFURBISHED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |