FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1894455 · Received November 9, 2010

Report

Report Number
1423500-2010-05541
Event Type
Malfunction
Date Received
November 9, 2010
Date of Event
September 30, 2010
Report Date
October 18, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL DEVICE WAS RETURNED AND EVALUATED. REVIEW OF THE DEVICE LOGS REVEALED AN INCREASED INTRAPERITONEAL VOLUME (IIPV) HAD OCCURRED WHEN THE USER DRAINED OUT 3304 ML, WHICH WAS GREATER THAN THE LARGEST PRESCRIBED FILL VOLUME OF 2000 ML AND MET IIPV CRITERIA. THE ASSIGNABLE CAUSE OF THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE: INSUFFICIENT DRAIN- FALSE EMPTY DETECT AND USE ERROR INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED TOO LOW (0ML) AND USE ERROR, TIDAL UF REMOVAL SET TOO LOW (0ML). A REVIEW OF THE PREVIOUS SERVICE RECORD SHOWS THE DEVICE PASSED ALL REQUIRED TESTS AND CALIBRATIONS PRIOR TO ITS RELEASE FROM THE (B)(4). NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE ISSUE OF THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WHICH OCCURRED ON (B)(6) 2010 DURING NITE 4 ULTRAFILTRATION, DRAIN VOLUME 3304ML. THIS EVENT MEETS THE OVERFILL CRITERIA

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1