FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 18944513 · Received March 20, 2024

Report

Report Number
2955842-2024-12656
Event Type
Malfunction
Date Received
March 20, 2024
Date of Event
February 27, 2024
Report Date
February 27, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE ERBE INTEGRATED ELECTRO SURGICAL UNIT (IESU) TO PERFORM FAILURE ANALYSIS. FAILURE ANALYSIS WAS CONFIRMED AND REPRODUCED. THE IESU WAS PLACED ON AN IN-HOUSE SYSTEM AND WAS RUN IN NORMAL MODE. ON STARTUP, THE IESU DISPLAYED A C-34 ERROR. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MALIGNANT HYSTERECTOMY SURGICAL PROCEDURE, CUSTOMER CONTACTED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) AND REPORTED THE ERBE INTEGRATED ELECTRO SURGICAL UNIT (IESU) FAULTED TOWARDS THE END OF THE CASE. THE SURGEON WAS ABLE TO FINISH CASE ROBOTICALLY. THE TSE REVIEWED LOGS AND FOUND ERROR C-34 AND M-11 ON ERBE IESU. PRIOR TO CONTACTING THE TSE, THE CUSTOMER POWER CYCLED THE ERBE IESU, BUT ISSUE PERSISTED. THE CUSTOMER CONNECTED ENERGY ACTIVATION CABLE TO SPARE GENERATOR TO RESOLVE THE REPORTED ISSUE. THE ISI REP. CONTACTED THE TSE AND REPORTED THE ERBE IESU HAD A C-00 ERROR. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PROCEDURE PERFORMED WAS HYSTERECTOMY. THE SYSTEM DID INITIALLY POWER ON WITHOUT ERRORS. THE PROCEDURE WAS COMPLETED WHEN THE ERBE IESU STOPPED WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2047491 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-40 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Female DA VINCI INSTRUMENTS AND ACCESSORIES.