ARCHITECT HAVAB IGM REAGENT
Report
- Report Number
- 2623532-2010-00079
- Event Type
- Malfunction
- Date Received
- November 9, 2010
- Report Date
- September 2, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS INT'L, LTD
- Product Code
- LOL
- PMA / PMN Number
- N/A
- Removal / Correction Number
- 2623532-1/4/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
(B)(4). EVALUATION: CROSS CONTAMINATION WAS IDENTIFIED AS A POTENTIAL CAUSE. CONCLUSION: INSUFFICIENT ACTIVE ANTIGEN IS BEING COATED ON THE MICROPARTICLE WITH THE CURRENT MICROPARTICLE MANUFACTURING PROCESS. AN ADVERSE TREND IN US CUSTOMER COMPLAINTS FOR ARCHITECT HAVAB-IGM ASSAY (LIST NUMBER 6L21) REGARDING HIGHER THAN USUAL GRAYZONE/REACTIVE (GZ/R) RESULTS RATE WAS DETECTED ON MARCH, 2009. THE INVESTIGATION REVEALED THE MOST PROBABLE CAUSE FOR THE INCREASE RATE OF GZ/R RESULTS IS THE HAV ANTIGEN CODE 26286 WHICH HAS A REDUCED POTENCY THAT AFFECTS THE ARCHITECT HAVAB-IGM PERFORMANCE. (B)(4). AS A RESULT, THE SIGNAL TO NOISE RATIO IS SHIFTED SPECIALLY IN THE NEGATIVE SAMPLES LEAVING THE ASSAY PRONE FOR FALSE GRAYZONE/REACTIVE RESULTS AND INCREASES ARCH HAVAB-IGM ASSAY SUSCEPTIBILITY TO REAGENT CROSS CONTAMINATION AND TEST SYSTEM VARIABILITY. AS A PREVENTIVE MEASURE TO PROTECT THE INTEGRITY OF TEST RESULTS, CUSTOMERS WERE INSTRUCTED BY A PRODUCT INFORMATION LETTER TO FOLLOW AN ALTERNATIVE PROCESSING METHOD BY SEGREGATING ALL HAVAB-IGM SAMPLES.
THE CUSTOMER STATED THERE WAS AN INCREASE IN (B)(6) RESULTS GENERATED ON THE ARCHITECT I2000SR ANALYZER. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED. THERE IS A POTENTIAL TO GENERATE (B)(6) OR (B)(6) RESULTS WHEN THE ARCHITECT AUSAB ASSAY PRECEDES (B)(6). A PRODUCT CORRECTION WAS ISSUED AND REPORTED UNDER 21CFR806 TO THE FDA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT HAVAB IGM REAGENT | FOR THE DETECTION OF IGM ANTIBODIES TO HEPATITIS A VIRUS | LOL | ABBOTT DIAGNOSTICS INT'L, LTD | 88077Q100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCH I2000SR, LN 3M74-98, (B)(4) |