FDA Adverse Event Injury Summary report: N

GORE® EXCLUDER® AAA ENDOPROSTHESIS

MDR report key: 18942758 · Received March 20, 2024

Report

Report Number
3013164176-2024-02019
Event Type
Injury
Date Received
March 20, 2024
Date of Event
February 22, 2024
Report Date
April 19, 2024
Manufacturer
W. L. GORE & ASSOCIATES, INC.
Product Code
MIH
PMA / PMN Number
P020004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B22 AND C20 ¿ PRODUCT HISTORY REVIEW COULD NOT BE PERFORMED AS THE SERIAL# IS NOT AVAILABLE. CODE C20 - THE DEVICE REMAINS IMPLANTED AND, THEREFORE, WAS NOT AVAILABLE FOR DIRECT ANALYSIS BY GORE. IT SHOULD BE NOTED THAT, PER THE GORE® EXCLUDER® AAA ENDOPROSTHESIS INSTRUCTIONS FOR USE (IFU), ADVERSE EVENTS THAT MAY OCCUR AND / OR REQUIRE INTERVENTION INCLUDE, BUT ARE NOT LIMITED TO, ENDOLEAK AND COMPONENT MIGRATION. PATIENT IMAGES WERE PROVIDED. IMAGINING ANALYSIS PROVIDED THE FOLLOWING: ¿ RECONSTRUCTION IMAGE SHOWS DEVICE DISCONNECTION, THEREBY, CONFIRMING THE TYPE III ENDOLEAK. ¿ THE DISTAL END OF THE CONTRALATERAL LEG USED AS THE IBE BRIDGING DEVICE HAS NO WALL APPOSITION WITHIN THE PROXIMAL IBE IN THE LCI. ¿ THE LENGTH FROM THE DISTAL END OF THE CONTRALATERAL LEG, WITHIN THE ABDOMINAL ANEURYSM SAC, TO THE PROXIMAL END OF THE IBE APPEARS TO BE ~6CM, BY OUTER CURVE LENGTH. ¿ THE DIAMETER AT THE DISTAL END OF THE CONTRALATERAL LEG IS 23.0MM. ¿ THE DIAMETERS WITHIN THE IBE TRUNK, IN THE LCI, RANGE FROM 23.0MM ¿ 24.3MM. W. L. GORE & ASSOCIATES, INC. (GORE) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY GORE, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. BLANK FIELDS PRESENT ON THIS REPORT INCLUDE REQUIRED FIELDS AND FIELDS DETERMINED TO BE NOT APPLICABLE. BLANK REQUIRED FIELDS INDICATE THAT THE INFORMATION WAS NOT PROVIDED, WAS DEEMED UNAVAILABLE OR WAS NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, GORE, OR ITS ASSOCIATES THAT THE DEVICE, GORE OR ITS ASSOCIATES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

FSA LATER REPORTED REINTERVENTION SURGERY WAS PERFORMED ON (B)(6) 2024. TWO CONTRALATERAL COMPONENTS WERE IMPLANTED TO EXTEND INTO THE IBE DEVICE.

Description of Event or Problem · 0

ON (B)(6) 2020, PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT AN AORTO-ILIAC ANEURYSM UTILIZING THE GORE® EXCLUDER® AAA ENDOPROSTHESIS AND GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS. ON 2/22/2024, FSA WAS NOTIFIED THAT THE CONTRALATERAL COMPONENT MIGRATED (6.5 CM) PROXIMALLY FROM THE ORIGINAL POSITION, DISCONNECTING FROM THE GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (ILIAC BRANCH COMPONENT), CAUSING A TYPE 3 ENDOLEAK. REINTERVENTION SURGERY IS SCHEDULED FOR (B)(6) 2024. PHYSICIAN DOES NOT KNOW WHAT CAUSED THIS ISSUE. FSA ALSO REPORTS HOSPITAL DID NOT SEND INITIAL DEVICE IMPLANT INFORMATION TO GORE, FURTHER PATIENT INFORMATION AND SERIAL# REMAINS UNKNOWN.

Description of Event or Problem · 0

FSA LATER REPORTED REINTERVENTION SURGERY WAS PERFORMED ON (B)(6) 2024. TWO CONTRALATERAL COMPONENTS WERE IMPLANTED TO EXTEND INTO THE IBE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2047372 GORE® EXCLUDER® AAA ENDOPROSTHESIS SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W. L. GORE & ASSOCIATES, INC.

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown