FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1894142 · Received November 4, 2010

Report

Report Number
2531779-2010-02088
Event Type
Malfunction
Date Received
November 4, 2010
Date of Event
October 5, 2010
Report Date
October 5, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND THE EVALUATION REVEALED THAT THE KEYPAD APPEARED TO BE SWOLLEN. ALL KEYS WERE INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. THE KEYPAD WAS REMOVED AND THE "UP" AND "CONTRAST" KEY CONTACTS WERE OBSERVED TO BE INVERTED AND MISALIGNED. ADDITIONALLY, THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE BUTTONS HAVE TO BE PRESSED MULTIPLE TIMES TO ELICIT A RESPONSE FROM THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA