FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 1894133 · Received November 4, 2010

Report

Report Number
2027969-2010-01879
Event Type
Malfunction
Date Received
November 4, 2010
Date of Event
October 14, 2010
Report Date
November 4, 2010
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K021923
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2010; INRATIO: 7.5; REFERENCE: 3.4; MEAN: 5.45; CONFIDENCE LIMITS: NA. THE MEAN IS >5.0 AND THE DIFFERENCE IS GREATER THAN 2.2. THESE RESULTS ARE CONSIDERED INACCURATE WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. ADDITIONAL INVESTIGATION IS REQUIRED. PREVIOUS INVESTIGATION OF STRIP LOT #232170 FROM ANOTHER CASE MET ACCURACY CRITERIA. NO FURTHER INVESTIGATION IS REQUIRED. THE ALLOWABLE DIFFERENCE BETWEEN THE INRATIO INR'S AND SYSMEX INR'S ARE CALCULATED. AT LEAST (B)(4) REPLICATES FOR BOTH SAMPLES FOR STRIP LOT #232170 ARE WITHIN THE ALLOWABLE BIAS (+ OR - 1.0). NO DISCREPANT RESULTS ARE PRODUCED ON RETURNED AND IN-HOUSE METERS. THESE MEET THE ABOVE CRITERIA, AND THEN NO FURTHER ACTION IS REQUIRED. ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON DID NOT MEET ACCURACY CRITERIA. ONE OUT OF TWO TOTAL NORMAL DONOR TEST RESULTS WERE OUT OF RANGE. COMPLAINT DID NOT CONTAIN SUFFICIENT INFORMATION FOR ROOT CAUSE ANALYSIS. NO PRODUCT WAS EXPECTED TO BE RETURNED. REFERENCE RETAIN STRIP TEST RESULT COMPARISON; WHICH MET ACCURACY CRITERIA. AS OF 11/04/2010, 27 DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #232170 YIELDING A COMPLAINT RATE OF 0.013%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2010; INRATIO: 7.5; LAB: 3.4.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100139 232170

Patients

Seq Age Sex Outcome Treatment
1