FDA Adverse Event Malfunction Summary report: N

SELECTSECURE MRI SURESCAN

MDR report key: 18940783 · Received March 20, 2024

Report

Report Number
2182208-2024-01178
Event Type
Malfunction
Date Received
March 20, 2024
Date of Event
January 1, 2023
Report Date
March 20, 2024
Manufacturer
MEDTRONIC, INC.
Product Code
NVN
PMA / PMN Number
P030036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. THIS EVENT OCCURRED OUTSIDE THE US. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. OF NOTE, MULTIPLE PATIENTS AND MULTIPLE MANUFACTURERS WERE NOTED IN THE ARTICLE; HOWEVER, A ONE-TO-ONE CORRELATION COULD NOT BE MADE WITH UNIQUE PRODUCT SERIAL/LOT NUMBERS. THE BASELINE GENDER/AGE CHARACTERISTICS IS UNKNOWN/57 YEARS OLD. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. REFERENCED ARTICLE: RIGHT SIDED APPROACH FOR LEFT BUNDLE BRANCH PACING USING LUMEN-LESS LEAD: TECHNICAL CONSIDERATIONS AND FOLLOW-UP OUTCOME. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY. 2023; 34:2613¿2616. DOI: 10.1111/JCE.16126 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

LITERATURE WAS REVIEWED REGARDING RIGHT SIDED APPROACH FOR LEFT BUNDLE BRANCH (LBB) PACING. THE AUTHORS DESCRIBED LEADS WHICH EXHIBITED INCREASES IN CAPTURE THRESHOLD, DECREASES IN PACING IMPEDANCE, AND LOSS OF LBB WITH INTACT SEPTAL CAPTURE DURING FOLLOW-UP. THE STATUS OF THE LEADS IS UNKNOWN. NO ADVERSE PATIENT EFFECTS OR ADDITIONAL PRODUCT PERFORMANCE ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2048276 SELECTSECURE MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) NVN MEDTRONIC, INC. 383069

Patients

Seq Age Sex Outcome Treatment
1 57 YR Unknown UNKNOWN COMPETITOR CIED AND LEADS