FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 1893981 · Received November 3, 2010

Report

Report Number
3015876-2010-01220
Event Type
Malfunction
Date Received
November 3, 2010
Date of Event
October 4, 2010
Report Date
October 4, 2010
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT POWER ON/OFF BY ITSELF FAILURE. PHYSIO DETERMINED THE CAUSE OF THE ISSUE TO BE A LOOSE BATTERY AND REPAIRED THE DEVICE. PROPER DEVICE OPERATION WAS THEN OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING BEFORE IT WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY POWER ON/OFF BY ITSELF. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA