FDA Adverse Event Malfunction Summary report: N

POSEY SITTER II

MDR report key: 1893954 · Received November 3, 2010

Report

Report Number
2020362-2010-00296
Event Type
Malfunction
Date Received
November 3, 2010
Report Date
July 22, 2010
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) (CUSTOMER DID NOT SPECIFY THE NATURE OF THE COMPLAINT). RESULTS - EVALUATION FOR THE RETURNED PRODUCT SHOWS THE ALARM TONE HAS STATIC NOISE WHEN ITS CHANGED AND WHEN PRESSURE IS APPLIED TO THE SENSOR PAD. THE ALARM PCB SHOWS CORROSION AND BATTERY LEAKAGE. THE UNIT BATTERY DOOR IS BROKEN AND SLIDES OUT FROM THE ALARM. POSEY INSTRUCTIONS RECOMMENDS: THE USE OF ALKALINE BATTERIES IN POSEY FALL ALARMS. DO NOT MIX OLD AND NEW BATTERIES, OR MIX DIFFERENT TYPES OF BATTERIES. THIS CAN CAUSE CORROSION AND LEAKAGE TO THE ALARM UNIT. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER DID NOT PROVIDE THE NATURE OF THE COMPLAINT. THERE WAS NO PATIENT INJURY OR INCIDENT REPORTED. EVALUATION FOR THE RETURNED PRODUCT SHOWS BATTERY LEAKAGE AND CORROSION.