FDA Adverse Event Malfunction Summary report: N

BARDEX® ALL-SILICONE FOLEY CATHETER

MDR report key: 18939515 · Received March 20, 2024

Report

Report Number
1018233-2024-01350
Event Type
Malfunction
Date Received
March 20, 2024
Date of Event
February 26, 2024
Report Date
February 28, 2024
Manufacturer
C.R. BARD INC. (COVINGTON) -1018233
Product Code
EZL
UDI-DI
00801741029752
PMA / PMN Number
K040504
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS INCONCLUSIVE BECAUSE NO SAMPLE WAS RETURNED. A POTENTIAL ROOT CAUSE FOR THIS FAILURE COULD BE "INCORRECT BALLOON DESIGN (BALLOON WALL THICKNESS EXCESSIVE) ¿. THE LOT NUMBER WAS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "PROPER TECHNIQUES FOR URINARY CATHETER INSERTION ¿ PERFORM HAND HYGIENE IMMEDIATELY BEFORE AND AFTER INSERTION ¿ INSERT URINARY CATHETERS USING ASEPTIC TECHNIQUE AND STERILE EQUIPMENT ¿ USE THE SMALLEST FOLEY CATHETER POSSIBLE, CONSISTENT WITH GOOD DRAINAGE ¿ DOCUMENT THE INDICATIONS FOR CATHETER INSERTION, DATE AND TIME OF CATHETER INSERTION, INDIVIDUAL WHO INSERTED CATHETER, AND DATE AND TIME OF CATHETER REMOVAL IN PATIENT RECORD PROPER TECHNIQUES FOR URINARY CATHETER MAINTENANCE ¿ SECURE THE FOLEY CATHETER, USE THE STATLOCK® FOLEY STABILIZATION DEVICE IF PROVIDED ¿ MAINTAIN A CLOSED DRAINAGE SYSTEM BY UTILIZING PRE-CONNECTED, SEALED CATHETER-TUBING JUNCTIONS ¿ MAINTAIN UNOBSTRUCTED URINE FLOW AND KEEP THE CATHETER AND COLLECTION TUBE FREE FROM KINKING ¿ KEEP THE COLLECTION BAG BELOW THE LEVEL OF THE BLADDER OR HIPS AT ALL TIMES ¿ EMPTY THE COLLECTION BAG REGULARLY (E.G., PRIOR TO TRANSPORT) USING A SEPARATE, CLEAN COLLECTION CONTAINER FOR EACH PATIENT ¿ ROUTINE HYGIENE (E.G., CLEANSING OF THE MEATAL SURFACE DURING DAILY BATHING OR SHOWERING) IS APPROPRIATE. ¿ LEAVE FOLEY CATHETER IN PLACE ONLY AS LONG AS NEEDED STERILE UNLESS PACKAGE IS OPENED OR DAMAGED, EXCEPT FOR ANY INDIVIDUALLY PACKAGED COMPONENTS WITHIN THE TRAY WHICH ARE NOT LABELED AS STERILE. THESE COMPONENTS ARE NOT TERMINALLY STERILIZED. CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. SINGLE USE ONLY. DO NOT RESTERILIZE. FOR UROLOGICAL USE ONLY. WARNING: ON CATHETER, DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE THE CATHETER AND MAY CAUSE BALLOON TO BURST. WARNING: AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH ACCEPTED MEDICAL PRACTICES AND APPLICABLE LOCAL, STATE AND FEDERAL LAWS AND REGULATIONS. VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE. IF PACKAGE IS OPENED OR IF ANY IMPERFECTION OR SURFACE DETERIORATION IS OBSERVED, DO NOT USE. PLEASE CONSULT PRODUCT LABEL AND INSERT FOR ANY INDICATIONS, CONTRAINDICATIONS, HAZARDS, WARNINGS, CAUTIONS AND DIRECTIONS FOR USE". H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.

Description of Event or Problem · 0

IT WAS REPORTED THAT INDWELLING BLADDER CATHETER REMOVED BY ICU REGISTERED NURSE, THEY DEFLATED CATHETER BALLOON, FELT RESISTANCE WHEN ATTEMPTING TO PULL OUT CATHETER AND UNABLE TO ASPIRATE ADDITIONAL FLUID FROM BALLOON. INTENSIVIST NOTIFIED THAT CATHETER WAS MEETING RESISTANCE AND UNABLE TO BE REMOVED SO THAT ADDITIONAL LUBRICANT APPLIED TO URETHRAL MEATUS, AND CATHETER WAS REMOVED. AFTER REMOVAL, CATHETER BALLOON VISIBLY SHAPED IRREGULARLY, LIKE A TULIP OR AN UPSIDE-DOWN FUNNEL. CATHETER HAD ORIGINALLY BEEN INSERTED ON 16FEB IN PACU.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1131393 BARDEX® ALL-SILICONE FOLEY CATHETER FOLEY CATHETER EZL C.R. BARD INC. (COVINGTON) -1018233 UNK 00801741029752

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other