PERSONAL CYCLER MACHINE 100 VOLTS
Report
- Report Number
- 1423500-2010-05494
- Event Type
- Malfunction
- Date Received
- November 9, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 15, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION RESULTS: THE TECHNICAL SERVICE CENTER RECEIVED THE ACTUAL SAMPLE FOR EVALUATION. THE ENGINEER PERFORMED AN EVALUATION BASED ON THE FAILURE ANALYSIS PROCEDURE. THE ENGINEER CONFIRMED THAT THERE WAS A PROBLEM ON THE POWER ENTRY MODULE. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE PROBLEM WITH A DEFECT OF THE ACCOMP BOARD AND HEATER PAN. THE POWER ENTRY MODULE WAS REPLACED AND THE INSTRUMENT THEN MET ALL SPECIFICATIONS. RENAL QUALITY ENGINEERING, ALONG WITH PLANT MANUFACTURING AND QUALITY PERSONNEL, WILL CONTINUE TO MONITOR PRODUCT LINES FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
(B)(4). THE DEVICE EVALUATION IS IN PROCESS. A FOLLOW-UP MEDWATCH WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
A (B)(6) PATIENT, AFTER POWERING ON THE HOMECHOICE (HC) DEVICE, REPORTED THAT THE POWER SHUT DOWN. IT WAS ALSO REPORTED THAT THE POWER BUTTON DID NOT WORK, THE NAVIGATIONAL PANEL WAS OFF AND AFTER PLUGGING IN AND UNPLUGGING THE DEVICE SEVERAL TIMES THE POWER ISSUE WAS NOT RESOLVED. THE TECHNICAL SERVICE CENTER RECEIVED THE ACTUAL SAMPLE FOR ROUTINE MAINTENANCE. AS PART OF THE MAINTENANCE EVALUATION, A TECHNICIAN CONFIRMED AND NOTED A BURNT PORTION OF THE POWER MODULE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERSONAL CYCLER MACHINE 100 VOLTS | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |