FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSETM
MDR report key: 1893877
·
Received November 1, 2010
Report
- Report Number
- 2531779-2010-02067
- Event Type
- Malfunction
- Date Received
- November 1, 2010
- Date of Event
- October 3, 2010
- Report Date
- October 3, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED SWOLLEN, THE PUMP WAS INTERMITTENTLY RESPONSIVE TO ALL KEYPAD BUTTONS AND REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE UP AND DOWN ARROW KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE BUTTON CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PUMP IS INTERMITTENTLY RESPONDING TO THE DOWN ARROW BUTTON. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO WATER AND THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSETM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |