FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSETM

MDR report key: 1893877 · Received November 1, 2010

Report

Report Number
2531779-2010-02067
Event Type
Malfunction
Date Received
November 1, 2010
Date of Event
October 3, 2010
Report Date
October 3, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED SWOLLEN, THE PUMP WAS INTERMITTENTLY RESPONSIVE TO ALL KEYPAD BUTTONS AND REQUIRED EXCESSIVE FORCE TO ACTIVATE, THE UP AND DOWN ARROW KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP IS INTERMITTENTLY RESPONDING TO THE DOWN ARROW BUTTON. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO WATER AND THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSETM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 18 YR