FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1893856 · Received November 1, 2010

Report

Report Number
2531779-2010-02064
Event Type
Malfunction
Date Received
November 1, 2010
Report Date
October 2, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED SWOLLEN, THE PUMP WAS INTERMITTENTLY RESPONSIVE TO ALL KEYPAD BUTTONS, THE UP KEY CONTACT WAS MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OK BUTTON HAS TO BE PRESSED VERY HARD FOR THE PUMP TO RESPOND. THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO WATER. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1