Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THAT THE POWER SUPPLY CONTACT WAS LOOSE ON THE UTV BOARD. HE RESEATED THIS CONNECTION AND THE NURSE CALL FUNCTIONED PROPERLY.
Description of Event or Problem · 1
THE TECHNICIAN DISCOVERED THE NURSE CALL WAS NOT WORKING WHILE HE WAS PERFORMING PREVENTIVE MAINTENANCE ON THE BED.