Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE POWER SUPPLY CONTACT WAS LOOSE ON THE UTV BOARD. HE RESEATED THIS CONNECTION AND THE NURSE CALL FUNCTIONED PROPERLY.

Description of Event or Problem · 1

THE TECHNICIAN DISCOVERED THE NURSE CALL WAS NOT WORKING WHILE HE WAS PERFORMING PREVENTIVE MAINTENANCE ON THE BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1