BANDER URETERAL DIVERSION STENT SET
Report
- Report Number
- 1820334-2024-00378
- Event Type
- Malfunction
- Date Received
- March 19, 2024
- Date of Event
- February 22, 2024
- Report Date
- September 17, 2024
- Manufacturer
- COOK INC
- Product Code
- FAD
- UDI-DI
- 00827002575837
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. G4: DEVICE IS NOT MARKETED IN THE UNITED STATES. THERE ARE SIMILAR DEVICES MARKETED IN THE UNITED STATES, FOR EXAMPLE CATALOG NUMBER 025807-S1. INFORMATION FOR 025807-S1: PMA/510(K) # (G4): = K181971. H3: DEVICE HAS BEEN RETURNED AND PRELIMINARY EVALUATION HAS BEEN PERFORMED, HOWEVER, OUR INVESTIGATION IS ONGOING AND DEVICE EVALUATION SUMMARY WILL BE INCLUDED IN OUR FOLLOW UP REPORT ONCE OUR INVESTIGATION HAS BEEN COMPLETED. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. CORRECTION: D4: MODEL # = GPN #. INVESTIGATION: EVALUATION. AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE UNOPENED PACKAGE OF A URETERAL DIVERSION STENT SET CONTAINED A FOREIGN SUBSTANCE. THE DEVICE WAS NOT USED. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT REPORTED. REVIEWS OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), AND QUALITY CONTROL (QC) PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. VISUAL INSPECTION OF THE RETURNED COMPLAINT DEVICE WAS ALSO CONDUCTED. ONE DEVICE UNOPENED WAS RETURNED FOR INVESTIGATION. FOREIGN MATTER COULD BE SEEN INSIDE THE STERILE BARRIER. ADDITIONALLY, A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. A REVIEW OF THE DEVICE MASTER RECORD (DMR) CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY THIS FAILURE MODE PRIOR TO DISTRIBUTION. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOUND NO RELATED NON-CONFORMANCES REPORTED FOR LOT. A COMPLAINT HISTORY DATABASE SEARCH SHOWED NO OTHER RELATED COMPLAINTS ASSOCIATED WITH THE FAILURE MODE FOR THE COMPLAINT DEVICE LOT. WHILE THE DEVICE WAS NOT MANUFACTURED TO CURRENT SPECIFICATION, BECAUSE THERE ARE NO RELATED NON-CONFORMANCES, ADEQUATE INSPECTION ACTIVITIES HAVE BEEN ESTABLISHED, THERE IS OBJECTIVE EVIDENCE THAT THE DHR WAS FULLY EXECUTED, AND NO OTHER LOT RELATED COMPLAINTS THAT HAVE BEEN RECEIVED FROM THE FIELD, IT WAS CONCLUDED THAT THERE IS NO EVIDENCE THAT NONCONFORMING PRODUCT EXISTS IN HOUSE OR IN FIELD. COOK ALSO REVIEWED PRODUCT LABELING; THE IFU SUPPLIED WITH THE DEVICE STATES THE FOLLOWING IN CONSIDERATION OF THE REPORTED FAILURE MODE: ¿HOW SUPPLIED DO NOT USE THE PRODUCT IF THERE IS DOUBT AS TO WHETHER THE PRODUCT IS STERILE.¿ THE COMPLAINT WAS CONFIRMED BASED ON EVALUATION OF THE RETURNED DEVICE. BASED UPON THE AVAILABLE INFORMATION AND RESULTS OF THE INVESTIGATION, COOK HAS CONCLUDED THAT THE CAUSE FOR THE COMPLAINT WAS A QUALITY CONTROL DEFICIENCY; EMPLOYEE AWARENESS TRAINING OF THE ISSUE WAS CONDUCTED. THE APPROPRIATE PERSONNEL HAVE BEEN NOTIFIED AND COOK WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. PER THE QUALITY ENGINEERING RISK ASSESSMENT, NO FURTHER ACTION IS REQUIRED. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.
AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE UNOPENED PACKAGE OF A URETERAL DIVERSION STENT SET CONTAINED A FOREIGN SUBSTANCE. THE DEVICE WAS NOT USED. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT REPORTED.
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1320529 | BANDER URETERAL DIVERSION STENT SET | FAD STENT, URETERAL | FAD | COOK INC | G57583 | 15803808 | 00827002575837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |