FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1893636
·
Received November 2, 2010
Report
- Report Number
- 2531779-2010-01993
- Event Type
- Malfunction
- Date Received
- November 2, 2010
- Date of Event
- September 29, 2010
- Report Date
- September 29, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE 'UP' AND 'OK' BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. THE 'UP AND 'OK' BUTTON CONTACTS WERE FOUND TO BE MISALIGNED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED.
Description of Event or Problem · 1
THE PT REPORTED THAT THE BUTTONS REQUIRE MULTIPLE PRESSES TO ELICIT A RESPONSE FROM THE PUMP. THE PT REPORTS NOT EXPOSING THE PUMP TO WATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |