FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1893633 · Received November 2, 2010

Report

Report Number
2531779-2010-02033
Event Type
Malfunction
Date Received
November 2, 2010
Date of Event
October 1, 2010
Report Date
October 1, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION, THE CONTRAST BUTTON RESPONDED TO HARD PRESSES AND THE OK AND DOWN BUTTONS WERE INTERMITTENTLY RESPONSIVE. INVESTIGATION FOUND ADHESIVE UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE OK AND DOWN ARROW BUTTONS ON THE KEYPAD ARE INTERMITTENTLY UNRESPONSIVE. SHE STATED THAT SHE FIRST NOTICED THE ISSUE TWO MONTHS AGO, IT IMPROVED FOR A SHORT TIME, AND THEN THE PROBLEM RETURNED FOUR DAYS AGO. SHE DENIED TRAUMA TO THE KEYPAD OR WATER EXPOSURE. SHE NOTED THAT SHE WEARS THE PUMP IN A POCKET, DOES NOT USE CLEANING PRODUCTS ON THE PUMP, AND DENIES PEELING/TEARING OF THE KEYPAD. THE PATIENT STATED THAT THE KEYPAD BUTTONS CLICK AND SPRING BACK NORMALLY EXCEPT FOR THE OK BUTTON. SHE REPORTED THAT THE PUMP IS USABLE WITH MONITORING OF BUTTON PRESSES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 22 YR