FDA Adverse Event Injury Summary report: N

DUROM US ACETABULAR COMPONENT 52/46 L

MDR report key: 1893612 · Received November 2, 2010

Report

Report Number
9613350-2010-00504
Event Type
Injury
Date Received
November 2, 2010
Date of Event
October 13, 2010
Manufacturer
ZIMMER GMBH
Product Code
KWA
Removal / Correction Number
9613350-07/15/2008-001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INFORMATION WAS FORWARDED FROM THE OWNER ESTABLISHMENT, ZIMMER INC., WHICH MARKETS THE DEVICES IN THE U.S. THE MANUFACTURER DID NOT RECEIVE EXPLANTED DEVICES, X-RAYS, OR OTHER SOURCE DOCUMENTS FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE EXPLANTED DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. WHILE A CAUSE FOR THIS SPECIFIC CLINICAL EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED, THE COMMON CLINICAL PRESENTATION AND THE DATE OF THE ORIGINAL IMPLANTATION SUGGEST THAT THE REVISION SURGERY IS RELATED TO THE ISSUES FOR WHICH ZIMMER IMPLEMENTED A CORRECTION IN JULY 2008. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE AND/OR THE DEVICE (S) BE RETURNED FOR EVALUATION AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER (B)(4) CONSIDERS THIS CASE CLOSED. (B)(4).

Description of Event or Problem · 1

IT IS REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON (B)(6) 2007. PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2010 DUE TO PAIN AND SUSPECTED CUP LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUROM US ACETABULAR COMPONENT 52/46 L DUROM ACETABULAR COMPONENT AND METASUL KWA ZIMMER GMBH NA 2349657

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization