FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 1893605 · Received November 1, 2010

Report

Report Number
1028232-2010-02306
Event Type
Injury
Date Received
November 1, 2010
Date of Event
August 24, 2010
Report Date
October 5, 2010
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE, BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFO THAT THIS RIGHT ATRIAL LEAD DISLODGED. AS A RESULT, THIS LEAD WAS SUCCESSFULLY REPOSITIONED. NO ADVERSE PT EFFECTS WERE NOTED. THIS LEAD REMAINS IMPLANTED. NO ADDITIONAL INFO IS AVAILABLE AT THIS TIME. SHOULD ADDITIONAL INFO BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEXTRUS 4136 PACEMAKER LEAD NVZ BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization