FDA Adverse Event Injury Summary report: N

OT ULTRASMART METER

MDR report key: 1893542 · Received November 8, 2010

Report

Report Number
2939301-2010-09705
Event Type
Injury
Date Received
November 8, 2010
Report Date
October 21, 2010
Manufacturer
LIFESCAN INC.
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT; 510(K) # IS K021819.

Description of Event or Problem · 1

ON (B)(6), 2010, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH ULTRASMART METER WAS READING INACCURATELY HIGH COMPARED TO HER FEELING/ NORMAL RESULT(S). THE FOLLOWING COMPLAINT WAS CLASSIFIED BASED ON INFORMATION OBTAINED FROM THE CUSTOMER SERVICE REPRESENTATIVE (CSR). THE PATIENT ALLEGED THAT THE ISSUE BEGAN ON THE MORNING OF (B)(6), 2010. THE PATIENT REPORTEDLY OBTAINED A BLOOD GLUCOSE RESULT OF "120 MG/DL" WITH THE SUBJECT METER. THE PATIENT INDICATED SHE MANAGES HER DIABETES WITH INSULIN (NO ADJUSTMENTS); HOWEVER, AFTER THE ALLEGED ISSUE OCCURRED, THE PATIENT DENIED TAKING ANY ACTION. APPROXIMATELY 20-30 MINUTES AFTER THE ALLEGED ISSUE BEGAN, THE PATIENT CLAIMED SHE DEVELOPED SYMPTOMS OF SWEATING, DIZZINESS, AND WAS ALMOST UNCONSCIOUS. THE PATIENT, HOWEVER, DENIED SHE RECEIVED ANY TREATMENT AS A RESULT OF THE REPORTED METER ISSUE. AT THE TIME OF TROUBLESHOOTING, THE CSR VERIFIED THE CORRECT UNIT OF MEASUREMENT ON THE SUBJECT METER. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF A SERIOUS INJURY AFTER THE ALLEGED METER ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRASMART METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3033084

Patients

Seq Age Sex Outcome Treatment
1 42 YR Life Threatening