FDA Adverse Event Malfunction Summary report: N

ARCHITECT C8000 SYSTEM

MDR report key: 1893503 · Received November 8, 2010

Report

Report Number
1628664-2010-00405
Event Type
Malfunction
Date Received
November 8, 2010
Date of Event
October 29, 2010
Report Date
November 1, 2010
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). REFER TO RESULTS OF INVESTIGATION BELOW. MULTIPLE PARTS REPLACED TO RESOLVE ISSUE. AN EVALUATION WAS PERFORMED BY REVIEWING THE COMPLAINT TEXT, INSTRUMENT LOGS, INSTRUMENT SERVICE HISTORY, AND OTHER CUSTOMER COMPLAINTS FOR SIMILAR ISSUES. A REVIEW OF COMPLAINTS FROM (B)(4) 2010 THROUGH (B)(4) 2010 DID NOT IDENTIFY AN ADVERSE TREND RELATED TO THE CUSTOMER'S ISSUE. ERRATIC RESULT RATES ARE BELOW THE INTERNAL RATE DOCUMENTED AT LAUNCH FOR THE ARCHITECT ANALYZER. BASED UPON THE AVAILABLE INFORMATION, A DEFINITIVE CAUSE FOR THE CUSTOMER'S ISSUE COULD NOT BE IDENTIFIED, HOWEVER SERVICE REPLACED SEVERAL PARTS ON THEIR INSTRUMENT, INCLUDING THE ICT MODULE, SYRINGES, VALVES AND TUBING, MIXER TIP, DRYER TIP, AND R1 AND R2 TUBING AND RESOLVED THE ISSUE. TROUBLESHOOTING AND DIAGNOSTICS IN THE ARCHITECT SYSTEM OPERATIONS MANUAL LISTS SEVERAL PROBABLE CAUSES AND CORRECTIVE ACTIONS FOR ERRATIC ICT RESULTS, INCLUDING SAMPLES, THE ICT MODULE, SYRINGES, CHECK VALVES AND TUBING, REAGENT PROBES, CUVETTE WASHER AND VARIOUS HARDWARE FAILURES. BASED ON THE DATA AVAILABLE AND THE RESULTS OF THE EVALUATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED AFTER INSTALLING A NEW ICT MODULE ON THE ARCHITECT C8000 ANALYZER, THERE HAVE BEEN OCCURRENCES OF HIGH CHLORIDE RESULTS. A PATIENT GENERATED AN INITIAL RESULT OF 127 MMOL/L BUT UPON REPEAT, THE RESULT WAS 103 MMOL/L. THE INITIAL RESULT WAS FLAGGED AS IT WAS HIGHER THAN PREVIOUS CHLORIDE RESULTS AND DID NOT ALIGN WITH OTHER RESULTS FROM THE SAME SAMPLE. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT C8000 SYSTEM AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT MANUFACTURING, INC.

Patients

Seq Age Sex Outcome Treatment
1 ICT CHLORIDE| ICT CHLORIDE