FDA Adverse Event Injury Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 18929032 · Received March 19, 2024

Report

Report Number
3008642652-2024-03022
Event Type
Injury
Date Received
March 19, 2024
Date of Event
March 12, 2024
Report Date
March 19, 2024
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BASED ON THE DATA AVAILABLE AT THIS TIME, THERE IS NO INDICATION OF A DEVICE MALFUNCTION CAUSING OR CONTRIBUTING TO THE INAPPROPRIATE TREATMENT. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. (B)(4). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF. THE LIFEVEST DETECTION ALGORITHM COMPLIES WITH IEC 60601-2-4 PERFORMANCE REQUIREMENTS FOR SENSITIVITY AND SPECIFICITY.

Description of Event or Problem · 0

THE PATIENT WAS INAPPROPRIATELY TREATED 1 TIME BY THE LIFEVEST. THE PATIENT WAS REPORTEDLY CONSCIOUS AT THE TIME OF THE EVENT. DOUBLE COUNTING OF THE ECG SIGNAL CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE PRESSED AFTER THE TREATMENT WAS DELIVERED. THE RESPONSE BUTTONS FUNCTIONED APPROPRIATELY. THE PATIENT REPORTED HE TRIED PRESSING THE RB¿S. NO INJURY WAS REPORTED FROM THE TREATMENT. THE PATIENT WENT TO THE HOSPITAL FOR EVALUATION. IT IS UNKNOWN IF THE PATIENT CONTINUED TO USE THE LIFEVEST. NO DEFICIENCIES WERE ALLEGED AGAINST THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2376204 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization