TALENT ABDOMINAL STENT GRAFT SYSTEM
MDR report key: 1892660
·
Received November 3, 2010
Report
- Report Number
- 2953200-2010-02140
- Event Type
- Injury
- Date Received
- November 3, 2010
- Date of Event
- October 8, 2010
- Report Date
- October 8, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). RESULTS - ENDOLEAK; RESULTS/CONCLUSIONS: AORTIC NECK ANGULATION WITH PROXIMAL CALCIFICATION.
Description of Event or Problem · 1
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 1 CM IN LENGTH, 23 MM IN DIAMETER, 40 DEGREE ANGLED AND WAS CALCIFIED PROXIMALLY. IT WAS REPORTED THAT AFTER THE BIFURCATED STENT GRAFT WAS IMPLANTED THERE WAS A PROXIMAL TYPE I ENDOLEAK PRESENT. A 28 MM DIAMETER ANEURX CUFF WAS IMPLANTED AND SUCCESSFULLY RESOLVED THE TYPE I ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00473942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Required Intervention |