FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 1892660 · Received November 3, 2010

Report

Report Number
2953200-2010-02140
Event Type
Injury
Date Received
November 3, 2010
Date of Event
October 8, 2010
Report Date
October 8, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS - ENDOLEAK; RESULTS/CONCLUSIONS: AORTIC NECK ANGULATION WITH PROXIMAL CALCIFICATION.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 1 CM IN LENGTH, 23 MM IN DIAMETER, 40 DEGREE ANGLED AND WAS CALCIFIED PROXIMALLY. IT WAS REPORTED THAT AFTER THE BIFURCATED STENT GRAFT WAS IMPLANTED THERE WAS A PROXIMAL TYPE I ENDOLEAK PRESENT. A 28 MM DIAMETER ANEURX CUFF WAS IMPLANTED AND SUCCESSFULLY RESOLVED THE TYPE I ENDOLEAK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00473942

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention