FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18926549 · Received March 18, 2024

Report

Report Number
2647876-2024-00057
Event Type
Malfunction
Date Received
March 18, 2024
Date of Event
December 23, 2023
Report Date
March 27, 2024
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
SEE H.10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY CATALOG: 442023. BATCH NO.: 3110054. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. NO TREND IDENTIFIED FOR BATCH. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

G.5 PMA / 510(K)# K113558, K222591. H.6 INVESTIGATION SUMMARY: CATALOG: 442023. BATCH NO.: 3110054. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. NO TREND IDENTIFIED FOR BATCH. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

REPORT 5 OF 8 IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THAT THERE WAS A FALSE POSITIVE FOR CANDIA TROPICAL ON PANEL BCID2. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

REPORT 5 OF 8. IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) THAT THERE WAS A FALSE POSITIVE FOR CANDIA TROPICAL ON PANEL BCID2. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2556428 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3110054 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown