MONARCH BRONCHOSCOPE
Report
- Report Number
- 3014447948-2024-00005
- Event Type
- Injury
- Date Received
- March 18, 2024
- Date of Event
- February 19, 2024
- Report Date
- March 18, 2024
- Manufacturer
- AURIS HEALTH, INC.
- Product Code
- QNW
- UDI-DI
- B634MBR000211A1
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE SUBJECT SCOPE WAS CONNECTED TO AN IN-HOUSE SYSTEM TO VERIFY ELECTRICAL FUNCTIONALITY OF THE CAMERA, LIGHT EMITTING DIODE (LED), AND ELECTROMAGNETIC SENSORS (EM). THE SCOPE TIP WAS MANUALLY ARTICULATED WHILE MONITORING THE RESISTANCE OF EACH SENSOR, AND THE CAMERA IMAGE TO CONFIRM THAT THERE WAS NO INTERMITTENT OPEN ON EITHER SENSOR OR CAMERA IMAGE DEGRADATION. THE RESULTS OBSERVED BOTH SENSORS TO BE WITHIN SPECIFICATION, AND THE CAMERA TO BE FUNCTIONAL WITH NO SIGNS OF IMAGE DEGRADATION. ADDITIONALLY, THE LED WAS NOT FUNCTIONING. THE SCOPE WAS CONNECTED TO THE IN-HOUSE SYSTEM FOR THIRTY MINUTES AND NO LED WAS OBSERVED. A CLOSER INSPECTION OF THE SCOPE HANDLE, SHAFT, CABLE, AND CONNECTOR OBSERVED NO EXTERNAL PHYSICAL DAMAGE THAT COULD CONTRIBUTE TO THE REPORTED ISSUE. NOTE: MANUFACTURER REPORT NUMBERS 3014447948-2024-00004 ARE FOR THE SAME PATIENT EVENT.
THE CLINICAL SPECIALIST REPORTED THAT DURING THE PROCEDURE, THE LIVE CAMERA VIEW ON THE SCOPE SUDDENLY BECAME STATIC AND FUZZY, LEADING TO A LOSS OF VIEW. THE USER SWITCHED SCOPES AND SUCCESSFULLY NAVIGATED TO THE LESION. HOWEVER, WHILE BIOPSYING, AN FAULT OCCURRED, INDICATING DISCONNECTION OF THE SCOPE AND DISABLING OF ROBOTIC FUNCTIONS. DESPITE ATTEMPTS, THE USER COULDN'T TURN THE CAMERA BACK ON OR CLEAR THE ERROR. THE CLINICAL SPECIALIST MENTIONED THAT THE ISSUE RECURRED WITH THE SECOND SCOPE DURING BIOPSYING. CONSEQUENTLY, THE PHYSICIAN DECIDED TO END THE PROCEDURE WITHOUT FURTHER BIOPSIES. THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT BECAUSE OF THE SCOPE ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186219 | MONARCH BRONCHOSCOPE | BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES | QNW | AURIS HEALTH, INC. | MBR-000211-A | B634MBR000211A1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |