FDA Adverse Event Injury Summary report: N

MONARCH BRONCHOSCOPE

MDR report key: 18921615 · Received March 18, 2024

Report

Report Number
3014447948-2024-00005
Event Type
Injury
Date Received
March 18, 2024
Date of Event
February 19, 2024
Report Date
March 18, 2024
Manufacturer
AURIS HEALTH, INC.
Product Code
QNW
UDI-DI
B634MBR000211A1
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SUBJECT SCOPE WAS CONNECTED TO AN IN-HOUSE SYSTEM TO VERIFY ELECTRICAL FUNCTIONALITY OF THE CAMERA, LIGHT EMITTING DIODE (LED), AND ELECTROMAGNETIC SENSORS (EM). THE SCOPE TIP WAS MANUALLY ARTICULATED WHILE MONITORING THE RESISTANCE OF EACH SENSOR, AND THE CAMERA IMAGE TO CONFIRM THAT THERE WAS NO INTERMITTENT OPEN ON EITHER SENSOR OR CAMERA IMAGE DEGRADATION. THE RESULTS OBSERVED BOTH SENSORS TO BE WITHIN SPECIFICATION, AND THE CAMERA TO BE FUNCTIONAL WITH NO SIGNS OF IMAGE DEGRADATION. ADDITIONALLY, THE LED WAS NOT FUNCTIONING. THE SCOPE WAS CONNECTED TO THE IN-HOUSE SYSTEM FOR THIRTY MINUTES AND NO LED WAS OBSERVED. A CLOSER INSPECTION OF THE SCOPE HANDLE, SHAFT, CABLE, AND CONNECTOR OBSERVED NO EXTERNAL PHYSICAL DAMAGE THAT COULD CONTRIBUTE TO THE REPORTED ISSUE. NOTE: MANUFACTURER REPORT NUMBERS 3014447948-2024-00004 ARE FOR THE SAME PATIENT EVENT.

Description of Event or Problem · 0

THE CLINICAL SPECIALIST REPORTED THAT DURING THE PROCEDURE, THE LIVE CAMERA VIEW ON THE SCOPE SUDDENLY BECAME STATIC AND FUZZY, LEADING TO A LOSS OF VIEW. THE USER SWITCHED SCOPES AND SUCCESSFULLY NAVIGATED TO THE LESION. HOWEVER, WHILE BIOPSYING, AN FAULT OCCURRED, INDICATING DISCONNECTION OF THE SCOPE AND DISABLING OF ROBOTIC FUNCTIONS. DESPITE ATTEMPTS, THE USER COULDN'T TURN THE CAMERA BACK ON OR CLEAR THE ERROR. THE CLINICAL SPECIALIST MENTIONED THAT THE ISSUE RECURRED WITH THE SECOND SCOPE DURING BIOPSYING. CONSEQUENTLY, THE PHYSICIAN DECIDED TO END THE PROCEDURE WITHOUT FURTHER BIOPSIES. THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT BECAUSE OF THE SCOPE ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186219 MONARCH BRONCHOSCOPE BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES QNW AURIS HEALTH, INC. MBR-000211-A B634MBR000211A1

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other