INTEGRATED APD SET W/CASSETTE3 PRONG
Report
- Report Number
- 1423500-2010-05374
- Event Type
- Malfunction
- Date Received
- November 5, 2010
- Date of Event
- October 13, 2010
- Report Date
- October 13, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF A SYSTEM ERROR (B)(4) (AIR IN SET). THE REPORTED CONDITION WAS NOT CONFIRMED DUE TO UNAVAILABLE SAMPLE. BASED ON THE INFORMATION OBTAINED DURING BAXTER'S INVESTIGATION, THE ROOT CAUSE WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE THERE WAS AN OPEN CLAMP ON AN UNUSED SUPPLY LINE. THE HOME PATIENT (HP) HAD 2 BAGS CONNECTED AND LEFT THE UNUSED LINE CLAMPS OPEN. A BATCH REVIEW WAS NOT PERFORMED BECAUSE THE LOT NUMBER WAS UNKNOWN. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THE SAMPLE WAS DISCARDED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.
THE HOME PATIENT(HP) CONTACTED BAXTER'S (B)(4) REGARDING A SYSTEM ERROR 2240 ALARM WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DWELL 4 OF 4. THE HP HAD 2 BAGS CONNECTED AND LEFT THE UNUSED LINE CLAMPS OPEN. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HP TO CLEAR THE ALARM. THE TSR REVIEWED THE SETUP WITH THE HP. THE TSR THEN REFERRED THE HP TO CALL THE NURSE TO INFORM. THE SOLUTION TO THIS ISSUE WAS PROVIDED OVER THE PHONE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |