AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2010-05370
- Event Type
- Malfunction
- Date Received
- November 5, 2010
- Date of Event
- October 13, 2010
- Report Date
- October 13, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR PROBLEM CODE CONNECTION ISSUE IS NOT CONFIRMED DUE TO LACK OF SAMPLE. ROOT CAUSE IS UNDETERMINED DUE TO LACK OF SAMPLE. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). THE REPORTED CONDITION WAS RESOLVED OVER THE PHONE; THE SAMPLE WAS DISCARDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
A HOME PATIENT (HP) CONTACTED (B)(4) REGARDING A CHECK LINES AND BAGS ALARM THAT OCCURRED ON THE HOMECHOICE (HC) UNIT DURING PRIME. THE HP STATED WHEN SHE WAS CHECKING THE LINES AND BAGS ONE OF THE SPIKES DISCONNECTED FROM THE BAG. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO START OVER USING NEW SUPPLIES. THE TSR FURTHER ASSISTED THE HP TO CLOSE ALL CLAMPS AND CYCLE POWER OFF/ON. THE HP CONFIRMED THE CASSETTE WAS REMOVED. THE HP CONFIRMED SHE WAS NOT CONNECTED TO THE HC: THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. ON (B)(6) 2010 PRODUCT SURVEILLANCE SPOKE WITH THE HP WHO STATED THIS EVENT WAS AN ISOLATED INCIDENT. THE HP CONFIRMED SHE USES A COMPACT EXCHANGE DEVICE (CXD) TO ASSIST HER WITH SPIKING HER BAGS. THE HP STATED THIS PARTICULAR NIGHT WAS THE ONLY TIME SHE HAS HAD ANY ISSUES. THE HP CONFIRMED SHE RESUMED THERAPY WITHOUT INCIDENT OR ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |