FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1891898 · Received November 5, 2010

Report

Report Number
1423500-2010-05364
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
May 17, 2010
Report Date
October 13, 2010
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER DISCONTINUED USE OF THE DEVICE FOR AUTOMATED PERITONEAL DIALYSIS THERAPY AND RETURNED IT TO THE (B)(4) FACILITY. THE DEVICE PASSED THE HOMECHOICE RETURN INSTRUMENT TEST / EVALUATION (RITE) ELECTRICAL TESTS PERFORMED BY THE PAL. REVIEW OF THE RETURNED DEVICE LOGS REVEALED AN ADDITIONAL DIFFICULTY OF AN INCREASED INTRAPERITONEAL VOLUME (IIPV) HAD OCCURRED ON (B)(6) 2010 DURING CYCLE 1 WHEN THE DEVICE USER DRAINED OUT 4673ML, WHICH WAS 187% OF THE FILL VOLUME OF 2500 ML AND MET IIPV CRITERIA. THE DEVICE WAS FOUND TO MEET FUNCTIONAL SPECIFICATIONS RELATIVE TO THE IIPV IDENTIFIED VIA DEVICE LOG REVIEW. THE ASSIGNABLE CAUSE OF THE IIPV FOUND IN THE LOGS WAS DETERMINED TO BE: INSUFFICIENT DRAIN; FALSE EMPTY DETECT AND USE ERROR, INAPPROPRIATE BYPASS OF THE LOW DRAIN VOLUME ALARM. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF THE HOME CHOICE DEVICE AN INCREASED INTRAPERITONEAL VOLUME EVENT WAS FOUND IN THE EVENT LOGS THAT OCCURRED ON (B)(6) 2010 DURING CYCLE 1, DRAIN VOLUME 4673 ML. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1