FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE

MDR report key: 18918671 · Received March 15, 2024

Report

Report Number
9617032-2024-00379
Event Type
Malfunction
Date Received
March 15, 2024
Date of Event
February 20, 2024
Report Date
April 17, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
FMI
UDI-DI
30382903688372
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

MATERIAL #: 368837. LOT/BATCH #: 3230300. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL DRAW TESTING AND NO ISSUES WERE OBSERVED RELATING TO SLEEVE LEAKAGE AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE SLEEVE LEAKAGE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE, THERE WAS SLEEVE LEAKAGE WHEN THE TUBE WAS INSERTED. ONE (1) DEVICE WAS AFFECTED. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE, THERE WAS SLEEVE LEAKAGE WHEN THE TUBE WAS INSERTED. ONE (1) DEVICE WAS AFFECTED. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
486853 BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE BLOOD SPECIMEN COLLECTION DEVICE FMI BECTON, DICKINSON AND COMPANY (BD) 3230300 30382903688372

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown