FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60 ARTICULATING

MDR report key: 1891683 · Received November 5, 2010

Report

Report Number
1527736-2010-00094
Event Type
Malfunction
Date Received
November 5, 2010
Date of Event
October 12, 2010
Report Date
October 14, 2010
Manufacturer
ETHICON ENDO SURGERY, INC (CINCINNATI)
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. THE EC60A DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH A GREEN CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE RETURNED CARTRIDGE RELOAD WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION BY RESETTING AND RELOADING IT INTO THE DEVICE. THE DEVICE ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED AS INTENDED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRECTOMY PROCEDURE, THE DEVICE WAS LOADED WITH A GREEN CARTRIDGE, CLOSED AND PUT THROUGH TROCAR. ONCE THROUGH THE TROCAR IT COULD NOT BE OPENED. THE JAWS WOULD NOT OPEN. THE DEVICE WAS PULLED OUT, ONCE OUT OF THE TROCAR, THE JAWS STILL COULD NOT BE OPENED. THIS WAS THE INITIAL FIRING ATTEMPT. A NEW DEVICE WAS USED TO COMPLETE. THERE WAS NO PATIENT IMPACT. THE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 ARTICULATING STAPLE, IMPLANTABLE GDW ETHICON ENDO SURGERY, INC (CINCINNATI) UNK

Patients

Seq Age Sex Outcome Treatment
1