FDA Adverse Event Injury Summary report: N

OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 10MM

MDR report key: 18915249 · Received March 15, 2024

Report

Report Number
0001038806-2024-00462
Event Type
Injury
Date Received
March 15, 2024
Date of Event
January 30, 2024
Report Date
June 19, 2024
Manufacturer
BIOMET 3I
Product Code
DZE
UDI-DI
00844868021957
PMA / PMN Number
K130949
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT (B)(4). A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO REPORT ADDITIONAL AND CORRECTED INFORMATION TO 0001038806-2024-00462. THE MANUFACTURING DATE HAS BEEN CORRECTED TO 6/1/20217. THE FOLLOWING SECTIONS ARE BEING REPORTED: H2: IF FOLLOW-UP, WHAT TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H4: DEVICE MANUFACTURER DATE. H6: TYPE OF INVESTIGATION. H10: ADDITIONAL NARRATIVES/DATA. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO MEDICAL CONDITIONS / PATIENT HABITS AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION AND NO ASSOCIATION BETWEEN THE DENTAL IMPLANT AND BONE LOSS WAS FOUND THAT CAN EXPLAIN THESE EVENTS. THEREFORE, THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS PROBLEMS REPORTED IN ASSOCIATION WITH IMPLANT FAILURE HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 13 WAS REMOVED DUE TO BONE LOSS. BONE LOSS AROUND IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS.

Description of Event or Problem · 0

NO ADDITIONAL OR CORRECTED INFORMATION TO REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1266186 OSSEOTITE TAPERED CERTAIN IMPLANT 4 X 10MM DENTAL IMPLANT DZE BIOMET 3I 2017050644 00844868021957

Patients

Seq Age Sex Outcome Treatment
1 45 YR Male Required Intervention