AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2010-00675
- Event Type
- Malfunction
- Date Received
- November 4, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 22, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). METHOD: THE RETURNED MR290 CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: A SMALL DENT WAS OBSERVED IN THE MIDDLE OF THE CHAMBER BASE, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100324. CONCLUSION: ALL MR290 CHAMBERS ARE VISUALLY INSPECTED FOR DAMAGE AND ARE PRESSURE TESTED FOR LEAKS PRIOR TO PACKING. ANY CHAMBER WHICH FAILS EITHER TEST IS DISCARDED. THIS SUGGESTS THAT THE CHAMBER HAD BEEN DAMAGED POST-PRODUCTION, POSSIBLY DURING TRANSPORT OR STORAGE. THE MR290 USER INSTRUCTIONS WARN THE USER NOT TO USE THE MR290 IF IT HAS BEEN DROPPED. OUR TRENDING AND MONITORING OF COSMETIC DEFECTS ON MR290 CHAMBERS HAS A RATE OF OCCURRENCE OF (B)(4).
(B)(4). THE COMPLAINT CHAMBER IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE BASE OF A NEW MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS FOUND TO BE DENTED WHEN IT WAS REMOVED FROM THE PACKAGING.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE BASE OF A NEW MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS FOUND TO BE DENTED WHEN IT WAS REMOVED FROM THE PACKAGING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 100324 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |