FDA Adverse Event Malfunction Summary report: N

AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 1891481 · Received November 4, 2010

Report

Report Number
9611451-2010-00672
Event Type
Malfunction
Date Received
November 4, 2010
Date of Event
October 12, 2010
Report Date
October 22, 2010
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT CHAMBER AND ASSOCIATED FEEDSET TUBING WERE VISUALLY INSPECTED FOR DAMAGE. RESULTS: A SMALL SPLIT WAS FOUND IN THE FEEDSET TUBING WHERE IT ATTACHES TO THE SPIKE. THE SURFACE OF THE SPLIT IS WAS ROUGH. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100127. CONCLUSION: THE SURFACE OF THE SPLIT TUBING WAS ROUGH, NOT SMOOTH. THIS SUGGESTS THAT THE TUBING HAD BEEN PULLED AWAY FROM THE DOME, POSSIBLY DUE TO THE FEEDSET BEING STRETCHED DURING THE WINDING PROCESS. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. OUR USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." OUR MONITORING AND TRENDING OF MR290 FEEDSET TUBING LEAKS HAS A RATE OF OCCURRENCE OF (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT CHAMBER IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT WATER LEAKED FROM BETWEEN THE BAG SPIKE AND WATER FEED TUBE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT WATER LEAKED FROM BETWEEN THE BAG SPIKE AND WATER FEED TUBE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 100127

Patients

Seq Age Sex Outcome Treatment
1