CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM
Report
- Report Number
- 2029046-2024-00855
- Event Type
- Malfunction
- Date Received
- March 15, 2024
- Date of Event
- February 19, 2024
- Report Date
- April 11, 2024
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- DYB
- UDI-DI
- 10846835016277
- PMA / PMN Number
- K170997
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE BWI PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY BIOSENSE WEBSTER, INC., OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, BIOSENSE WEBSTER, INC. OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. MANUFACTURER'S REFERENCE NUMBER: (B)(4).
ON 19-MAR-2024, THE PRODUCT INVESTIGATION WAS COMPLETED. IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT-SIDED ATRIAL TACHYCARDIA ABLATION PROCEDURE WITH A CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM AND THE MEDICAL TEAM NOTED A VALVE LEAKAGE. THERE WERE NO BROKEN PARTS VISIBLE, BUT DROPLETS AT THE HEMOSTATIC VALVE CLEARLY VISIBLE. THE VALVE, BRIM CAP, AND HUB WERE NOT DISLODGED OR OTHERWISE DAMAGED. NO SURGICAL DELAY ENSUED. THE PROCEDURE WAS COMPLETED. NO PATIENT CONSEQUENCES WERE REPORTED. DEVICE EVALUATION DETAILS: A PICTURE WAS RECEIVED FOR EVALUATION FOLLOWING BIOSENSE WEBSTER'S PROCEDURES. ACCORDING TO PICTURES PROVIDED BY THE CUSTOMER, THE PHOTO DOES NOT PROVIDED SUFFICIENT INFORMATION RELATED TO THE REPORTED VALVE LEAK ISSUE BY THE CUSTOMER, AND THEREFORE NO RESULTS CAN BE OBTAINED FROM IT. THE PRODUCT WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED THAT THE HEMOSTATIC VALVE WAS BROKEN IN THE STAR SECTION. A MICROSCOPIC EXAMINATION OF THE HEMOSTATIC VALVE SURFACE SHOWED STRESS MARKS ON THE DAMAGE. THE BROKEN CONDITION COULD BE RELATED TO THE INCORRECT INSERTION OF THE DILATOR INTO THE SHEATH; THE STRESS MARKS AND PHYSICAL DAMAGE OBSERVED SUGGEST THAT EXCESSIVE FORCE OR MANIPULATION WAS APPLIED; HOWEVER, THIS COULD NOT BE CONCLUSIVELY DETERMINED. DEVICE HISTORY RECORD WAS PERFORMED FOR THE FINISHED DEVICE 60000298 NUMBER, AND NO INTERNAL ACTIONS RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. THE ISSUE REPORTED BY THE CUSTOMER WAS CONFIRMED. THE OPTIMAL DEVICE PERFORMANCE GUIDE (ODP) CONTAINS THE FOLLOWING CAUTION: ALWAYS INSERT A DILATOR STRAIGHT INTO THE CENTER OF THE SHEATH'S VALVE TO PREVENT DAMAGE TO THE VALVE. DO NOT INSERT A DILATOR AT AN ANGLE, AS DAMAGE TO THE SHEATH VALVE MAY OCCUR. AS PART OF THE QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT-SIDED ATRIAL TACHYCARDIA ABLATION PROCEDURE WITH A CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM AND THE MEDICAL TEAM NOTED A VALVE LEAKAGE. THERE WERE NO BROKEN PARTS VISIBLE, BUT DROPLETS AT THE HEMOSTATIC VALVE CLEARLY VISIBLE. THE VALVE, BRIM CAP, AND HUB WERE NOT DISLODGED OR OTHERWISE DAMAGED. NO SURGICAL DELAY ENSUED. THE PROCEDURE WAS COMPLETED. NO PATIENT CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1320112 | CARTO VIZIGO¿ 8.5F BI-DIRECTIONAL GUIDING SHEATH - MEDIUM | INTRODUCER, CATHETER | DYB | BIOSENSE WEBSTER INC | 60000298 | 10846835016277 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |