FDA Adverse Event Malfunction Summary report: N

EVIS LUCERA ELITE VIDEO SYSTEM CENTER

MDR report key: 18914344 · Received March 15, 2024

Report

Report Number
3002808148-2024-02472
Event Type
Malfunction
Date Received
March 15, 2024
Date of Event
March 4, 2024
Report Date
April 12, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S NO VIDEO MALFUNCTION WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 8 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS PRESUMED THAT DUE TO PROLONGED USE, A SHORT CIRCUIT ON PRINTED CIRCUIT BOARD (DX BOARD) CAUSED NO VIDEO OUTPUT FROM SERIAL DIGITAL INTERFACE (SDI) OUTPUT TERMINAL. HOWEVER, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED AND THE EVALUATION FOUND NO VIDEO FROM THE SERIAL DIGITAL INTERFACE OUTPUT TERMINAL DUE TO HAVING NO IMAGE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD NO IMAGE FROM THE SERIAL DIGITAL INTERFACE OUTPUT TERMINAL. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED BY OUTPUTTING FROM A DIFFERENT TERMINAL; USING THE SAME SET OF EQUIPMENT. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH. ADDITIONAL DETAILS RELATING TO THE PATIENT AND THE EVENT HAVE BEEN REQUESTED, BUT NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1411451 EVIS LUCERA ELITE VIDEO SYSTEM CENTER VIDEO SYSTEM CENTER FET SHIRAKAWA OLYMPUS CO., LTD. CV-290

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown LIGHT SOURCE